Article ID Journal Published Year Pages File Type
1245161 Talanta 2007 5 Pages PDF
Abstract

A simple and rapid method using micellar electrokinetic capillary chromatography (MEKC) was developed for the separation and determination of acrylamide in potato chips at low levels for the first time. The experimental conditions for the separation and quantification of acrylamide were optimized at first. The optimized conditions were: 50 mmol/L Na2B4O7 and 40 mmol/L SDS at pH 10.0, 12 kV applied voltage, 76 cm total length (67 cm effective length) and 75 μm i.d. capillary, 198 nm wavelength, 15 cm high 25 s hydrodynamics sample injection, 20 °C air-cooling. The linear response of acrylamide concentration ranges from 0.5 to 100 μg/mL with high correlation coefficient (r = 0.9986, n = 9). The LOD and LOQ were estimated to be 0.1 and 0.33 μg/mL based on S/N = 3 and 10. The precision values (expressed as R.S.D.) of intra- and inter-day were 0.86–4.35% and 2.61–9.65%, respectively. Recoveries spiked at levels 2, 20, 60 μg/mL ranged between 90.86% and 99.6% with R.S.D. less than 6.5%. Finally, the developed method has been applied to the analysis of real samples and has achieved satisfactory results. All of these indicated that it was a reliable method for the quantification of acrylamide in potato chips.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,