Article ID Journal Published Year Pages File Type
1245801 Talanta 2006 5 Pages PDF
Abstract

A non-alcoholic coulometric reagent based on N-methylformamide (NMF) was shown to eliminate the severe interference effect caused by the alcohol component of the conventional Karl Fischer (KF) reagent on the battery electrolyte lithium bis(oxalato)borate (LiBOB). For sample amounts up to 240 μg of water, the stoichiometry of the KF reaction deviated only slightly from the ideal 1:1 ratio for the best reagent composition. Both solid and dissolved (in acetonitrile, tetrahydrofuran (THF), and ethylene carbonate/ethyl methyl carbonate) LiBOB were titrated successfully using a Metrohm 756 KF Coulometer with a diaphragm cell. The detection limit was estimated to be 0.5–1 μg of water using 100 ml of reagent in this system.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,