Article ID Journal Published Year Pages File Type
1245861 Talanta 2007 7 Pages PDF
Abstract

A novel, fast and simple liquid chromatographic method was developed and validated for the simultaneous determination of rosiglitazone and metformin in human plasma. The analysis was performed on a phenyl column (250 mm × 4.6 mm i.d., 5 μm) using a gradient method starting with mobile phase composed of acetonitrile:5 mM acetate buffer pH 5.5 (75:25, v/v). The flow rate was 1 mL min−1. UV detection was performed at 245 nm and verapamil was used as internal standard. The total run time was less than 10 min. Sample preparation included a simple protein precipitation step with acetonitrile. Validation experiments were performed to demonstrate stability, specificity, sensitivity, linearity, accuracy, precision and robustness. The limit of quantification was 100 ng mL−1 for rosiglitazone and 250 ng mL−1 for metformin. The extraction recoveries were 100.02–105.0% for rosiglitazone and 105.64–103.88% for metformin. The method was applied with success to plasma samples obtained from diabetic patients undergoing treatment with rosiglitazone and metformin.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,