Article ID Journal Published Year Pages File Type
1246124 Talanta 2007 6 Pages PDF
Abstract

In-channel indirect amperometric detection mode for microchip capillary electrophoresis with positive separation electric field is successfully applied to some heavy metal ions. The influences of separation voltage, detection potential, the concentration and pH value of running buffer on the response of the detector have been investigated. An optimized condition of 1200 V separation voltage, −0.1 V detection potential, 20 mM (pH 4.46) running buffer of 2-(N-morpholino)ethanesulfonic acid (MES) + l-histidine (l-His) was selected. The results clearly showed that Pb2+, Cd2+, and Cu2+ were efficiently separated within 80 s in a 3.7 cm long native separation PDMS/PDMS channel and successfully detected at a single carbon fibre electrode. The theoretical plate numbers of Pb2+, Cd2+, and Cu2+ were 1.2 × 105, 2.5 × 105, and 1.9 × 105 m−1, respectively. The detection limits for Pb2+, Cd2+, and Cu2+ were 1.3, 3.3 and 7.4 μM (S/N = 3).

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,