Article ID Journal Published Year Pages File Type
1246561 Talanta 2006 5 Pages PDF
Abstract

A hydrophilic transparent triacetyl cellulose membrane was adopted as a pH optode by immobilizing highly selective and sensitive Nile blue indicator on the membrane. Contrary to the common procedure for determinations using optodes, in which a steady state response is measured, a new approach is introduced in which the dynamic response of the optode is used as the analytical signal. While in common procedures, pH optodes exhibit limited linear dynamic range (often 2–4 pH units only), it is shown that in a time-based flow method, an optode with only one acid-base indicator can be used for measurement in the pH range of 0–10. The procedure is simple, inexpensive and rapid.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry