Article ID Journal Published Year Pages File Type
1246780 Talanta 2009 6 Pages PDF
Abstract

An automatic sampling device, i.e., process sampling module (PSM), connected with a purge and trap–GC–FID system has been developed for real-time monitoring of VOCs in wastewater. The system was designed to simultaneously monitor 17 compounds, including one polar compound, i.e., acetone, and 16 non-polar compounds. The trapping tube is packed with two adsorbents, Carbopack B and Carbosieve III, to trap target compounds. For the purpose of in situ monitoring, the flush valve of the sampling tube is composed of two two-way valves and a time controller to prevent absorption interference of the residue. The optimal conditions for the analytical system include a 12 min purge time at a temperature of 60 °C, and 4 min of desorption time with a desorption temperature of 260 °C. Good chromatograms have been obtained with the analytical system even if a cryogenic device and de-misting were not used. The relative standards deviation (RSD) of the system is between 2% and 13.4%, and accuracies between 0.3 and 23.5% have been achieved. The detection limits of the method range from 0.32 to 2.39 ppb. In this system, the four parts, i.e., PSM, P&T, GC, and FID, were simple, reliable and rugged. Also, the interface of these four parts was simple and dependable.

Keywords
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , ,