Article ID Journal Published Year Pages File Type
1247297 Talanta 2006 7 Pages PDF
Abstract

A technique for the analysis of organophosphorus pesticides (ethoprop, diazinon, disulfoton, fenthion) in aqueous sample using liquid-phase microextraction (LPME), coupled with gas chromatography–mass spectrometry (GC–MS) was developed. A small section of a hollow fiber inserted into the needle of GC syringe and filled with the 3.5 μl of organic solvent was used to extract pesticides from a 20 ml aqueous sample. The limits of detection (LOD) with the selected ion monitoring (SIM) mode varied from 0.2 to 0.006 μg/l. The calibration curves were linear over three orders of magnitude with R2 ≥ 0.996. The relative standard deviations of the analysis (inter- and intra-day) were 5–8%, and the relative recoveries from the lake water sample were greater than 83%. The results were compared with results obtained using solid-phase microextraction (SPME/GC/MS).

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,