Article ID Journal Published Year Pages File Type
1247373 Talanta 2006 6 Pages PDF
Abstract

Uranium single particle analysis has been performed by inductively coupled plasma-mass spectrometry (ICP-MS) and the performances are compared with that provided by scanning electron microsopy and single particle counting. The transient signal induced by the flash of ions due to the ionisation of an uranium colloidal particle in the plasma torch can be detected and measured for selected uranium ion masses (238U+, 235U+ or 254[238U16O]+) by the mass spectrometer. The signals recorded via time scanning are analysed as a function of particle size or fraction of the studied element or isotope in the colloid phase. The frequency of the flashes is directly proportional to the concentration of particles in the colloidal suspension. The feasibility tests were performed on uranium dioxide particles. The study also describes the experimental conditions and the choice of mass to detect uranium colloids in a single particle analysis mode.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry