Article ID Journal Published Year Pages File Type
1261162 Journal of the Association of Arab Universities for Basic and Applied Sciences 2015 6 Pages PDF
Abstract

A viable cost-effective and isocratic approach employing C-18 column (250 mm × 4.6 mm, 5 μm) based HPLC has been utilized to separate and estimate the drug, tranexamic acid in pharmaceutical formulations. Tranexamic acid contains no π-electrons to act as fluorophore or chromophore hence pre-column derivatization was performed with benzene sulfonyl chloride in aqueous medium at room temperature. The derivatized drug was then estimated using C-18 column by exploiting a 25:75 (v/v) solvent mixture of acetonitrile and 0.1 M ammonium acetate (pH 5.0) as the mobile phase. The flow rate of mobile phase was 1 mL/min and detection was performed at a wavelength of 232 nm using UV detector. Retention time of tranexamic acid was 4.42 min. The method followed linear regression equation in the concentration range of 1–100 μg/mL with co-efficient of determination equal to 0.9994. The limit of detection and limit of quantitation were 0.3 and 1 μg/mL, respectively. The relative standard deviation and recovery ranges for tranexamic acid were found to be 0.11–2.47% and 97.60–103.25%, respectively. The suggested method is very sensitive and may have the potential to be used for tranexamic acid detection in medicinal formulations.

Related Topics
Physical Sciences and Engineering Chemistry Chemistry (General)
Authors
, , , , , ,