| Article ID | Journal | Published Year | Pages | File Type |
|---|---|---|---|---|
| 1427874 | Materials Science and Engineering: C | 2016 | 6 Pages |
•A sensitive method to determine cTnI cardiac biomarker by EIS method was introduced.•Porous graphene oxide was suggested as an interesting scaffold for anti-cTnI antibody immobilization.•Good selectivity and specificity were observed for cTnI electrochemical biosensor based on Ag-Ab interaction.•A wide linear range of response to cTnI protein concentration with an excellent LOD was achieved.
Herein, we report the fabrication of a novel label-free impedimetric biosensor employing porous graphene oxide (PrGO) nanostructures for the specific detection of cardiac troponin-I (cTnI) to establish the myocardial infarction (MI). This nano-immunosensor demonstrates an outstanding selectivity and high sensitivity towards the human-cTnI analyte. An excellent detection limit of 0.07 ng mL− 1 and dynamic linear range of 0.1–10 ng mL− 1 were calculated for anti-cTnI/PrGO/GC. Finally, this biosensor was employed to check the concentration of the MI biomarker in real clinical samples and the results are in good agreement with standard enzyme-linked fluorescence assay (ELFA) method.
