Abstract

A simple and stable method based on a single-element interferometer for accurately measuring refractive index of transparent liquids was demonstrated. The refractive index is measured by rotating a rectangular optical glass cell which contains sample liquid and air simultaneously, and by calculating interference fringe shift number which is detected from an interferogram. This method was successfully used to measure the refractive indices of various transparent liquids including distilled water, ethanol and NaCl–water and ethanol–water solutions at various concentrations. The temperature- dependent refractive index of distilled water was also measured. Furthermore, our method is simple to implement, vibration insensitive, and of high accuracy up to 10−4.