| Article ID | Journal | Published Year | Pages | File Type | 
|---|---|---|---|---|
| 1534854 | Optics Communications | 2014 | 4 Pages | 
Abstract
												This paper describes an experimental setup for the spectral calibration of bolometric detectors used in radioastronomy. The system is composed of a Martin-Puplett interferometer with two identical artificial blackbody sources operating in the vacuum mode at 77 K and 300 K simultaneously. One source is integrated into a liquid nitrogen cryostat, and the other one into a vacuum chamber at room temperature. The sources were designed with a combination of conical with cylindrical geometries thus forming an orthogonal configuration to match the internal optics of the interfermometer. With a simple mathematical model we estimated emissivities of ε~0.995 for each source.
											Related Topics
												
													Physical Sciences and Engineering
													Materials Science
													Electronic, Optical and Magnetic Materials
												
											Authors
												Angel Colin, 
											