Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1599163 | Acta Metallurgica Sinica (English Letters) | 2009 | 5 Pages |
Abstract
A novel method was reported to measure the remnant magnetic field in Lorentz mode in a FEI Tecnai F20 transmission electron microscope equipped with a Lorentz lens. The movement of the circle Bloch line of the cross-tie wall in Permalloy is used to measure the remnant magnetic field by tilting the specimen and adjusting the objective lens current. The remnant magnetic field is estimated to be about 17 Oe, in a direction opposite to that of the objective lens magnetic field. The remnant magnetic field can be compensated by adjusting the value of the objective lens current.
Keywords
Related Topics
Physical Sciences and Engineering
Materials Science
Metals and Alloys
Authors
Haihua LIU, Xiaokun DUAN, Renchao CHE, Zhifeng WANG, Xiaofeng DUAN,