Article ID Journal Published Year Pages File Type
1678049 Ultramicroscopy 2010 4 Pages PDF
Abstract

We developed a new electron optical system with three dodecapoles to compensate for spherical aberration and six-fold astigmatism, which generally remains in a two-hexapole type corrector. In this study, we applied the corrector for image-forming system in transmission electron microscope. Compensation for higher-order aberration was demonstrated through a diffractogram tableau using a triple three-fold astigmatism field system, which was then compared with a double hexapole field system. Using this electron optical system, six-fold astigmatism was measured to be less than 0.1 mm at an acceleration voltage of 60 kV, showing that the system successfully compensated for six-fold astigmatism.

Related Topics
Physical Sciences and Engineering Materials Science Nanotechnology