Article ID Journal Published Year Pages File Type
1678125 Ultramicroscopy 2010 9 Pages PDF
Abstract

A software-based method for collecting precession electron diffraction (PED) patterns is described. The PED patterns are obtained on a computer controlled transmission electron microscope. A series of electron diffraction (ED) patterns are collected as still ED frames at equal intervals, while the electron beam is precessed by one period (360°) around the optical axis. A PED pattern is obtained by combining the different ED frames, which resembles the sampling of a conventional PED pattern. Since intermediate ED frames are collected, it is possible to perform different post-processing strategies on the ED data. This can be used for geometric corrections to obtain accurate integrated intensities. The alignments and data collection are fully automated and controlled by software. The data quality is comparable to what can be achieved using specialized hardware for precession. The PED data can be used for structure solution and refinement with reasonably good R-values.

Research Highlights► Precession electron diffraction without an extra hardware. ► Computer control of the electron beam in TEM. ► Each different orientation saved as a separate file, allowing different processing. ► Geometric (Lorentz) factors corrected for. ► Electron precession without double spots.

Related Topics
Physical Sciences and Engineering Materials Science Nanotechnology