Article ID Journal Published Year Pages File Type
1680872 Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 2014 5 Pages PDF
Abstract

A sample of known elemental concentrations was activated in the bremsstrahlung photon beam which was created by a pulsed electron LINAC. Several procedures of photon activation analysis, including those applied with/without reference material and with/without photon flux monitor, were conducted to make a comparison of their precision and accuracy in practice. Experimental results have indicated that: (1) relative procedures usually produce better outcome despite that the absolute measurement is straightforward and eliminate the assistance of reference materials; (2) among relative procedures, the method with internal flux monitor yields higher quality of the analytical results. In the article, the pros and cons of each procedure are discussed as well.

Related Topics
Physical Sciences and Engineering Materials Science Surfaces, Coatings and Films
Authors
, , , , ,