Article ID Journal Published Year Pages File Type
1682953 Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 2011 4 Pages PDF
Abstract
The fast control of the beam spill extracted from a synchrotron is a key function for the spot scanning irradiation in cancer therapy application. The authors propose an extraction method which uses the quadruple field of fast response, as well as the RF-knockout. A Panofsky magnet was developed as a quadruple magnet, with a frequency response of around 10 kHz. The Panofsky magnet has a rectangular beam aperture and plate coils attached to the pole face. A model magnet has been manufactured with ferrite, and static and dynamic magnetic fields were measured. From the measurement we observed that the effects of eddy current in the plate coils were large and the uniformity of the magnetic field gradient in the beam aperture was worse than ±5% with a plate thickness of 0.02 cm and a frequency of current of 10 kHz. For the future, in a detailed design the eddy current effects have to be taken into account.
Related Topics
Physical Sciences and Engineering Materials Science Surfaces, Coatings and Films
Authors
, ,