Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1683224 | Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms | 2010 | 5 Pages |
Abstract
A fast and simple method for the determination of the efficiency coefficient (η) of bare CR-39 detector is presented and discussed. The efficiency coefficient of bare CR-39 detector is then calculated by different ways and the obtained values are found to be comparable to each other. The average value of η of bare CR-39 is found to be 0.20 ± 0.01 tracks cmâ2 dayâ1 per Bq mâ3.
Keywords
Related Topics
Physical Sciences and Engineering
Materials Science
Surfaces, Coatings and Films
Authors
Nidal Dwaikat, Mousa El-hasan, Masto Sueyasu, Wataru Kada, Fuminobu Sato, Yushi Kato, G. Saffarini, Toshiyuki Iida,