Abstract

A time-of-flight ERDA (TOF-ERDA) measurement system has been developed for the analysis of light elements. He ions are used for the incident beam, and recoil light ions are detected with the system. The system consists of a time detector and a silicon detector, and energy and velocity of recoil ion are measured simultaneously. The depth resolution of 21.6 ± 2.2 nm (FWHM) has been obtained by an ERDA measurement of a thin carbon layer onto a silicon wafer using a 5.7 MeV He beam. The mass resolution is better than 1 for elements up to oxygen. Maximum detectable depth of carbon in a PET film is about 650 nm. An ERDA measurement of implanted carbon in a silicon wafer has been demonstrated.

Related Topics
Physical Sciences and Engineering Materials Science Surfaces, Coatings and Films