Article ID Journal Published Year Pages File Type
1685674 Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 2007 5 Pages PDF
Abstract

We have built a velocity map imaging (VMI) spectrometer optimized for angle-resolved photoionization experiments with synchrotron radiation (SR) in the VUV and soft X-ray range. The spectrometer is equipped with four electrostatic lenses that focus the charged photoionization products onto a position-sensitive multi-hit delay-line anode. The use of two additional electrostatic lens elements as compared to the standard design of Eppink and Parker [A.T.J.B. Eppink, D.H. Parker, Rev. Sci. Instr. 68 (1997) 3477] provides better focusing of an extended interaction region, which is crucial for most SR applications. Furthermore, the apparatus is equipped with a second micro-channel plate detector opposite to the VMI spectrometer, enabling electron–ion coincidence experiments and thereby mass-resolved ion spectroscopy independent of the time structure of the synchrotron radiation. First results for the photofragmentation of CO2 molecules are presented.

Related Topics
Physical Sciences and Engineering Materials Science Surfaces, Coatings and Films