Article ID Journal Published Year Pages File Type
1685806 Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 2007 9 Pages PDF
Abstract

In a short time Be, C, Al, Cl, Ca and I accelerator mass spectrometry (AMS) have been established on the National Electrostatics Corporation (NEC) 5 MV pelletron system at the Scottish Universities Environmental Research Centre (SUERC). While summarising the present performance of the system, this report will focus on the details of ion detection, which sample materials are used and the analytical procedures employed for each individual species during routine analysis.All rare isotope detection is with a single flexible detector and ion event analysis system, but switching of analysed species typically requires a detector reconfiguration. Configurations for routine 10Be, 14C, 26Al, 36Cl, 41Ca and 129I detection have been established and will be presented here. Notably, there has proven to be sufficient suppression of the isobaric interferences of 36Cl and 41Ca in the 5+ charge state using an argon gas stripper at a terminal voltage of 5.0 MV to allow for routine analysis of these isotopes.

Related Topics
Physical Sciences and Engineering Materials Science Surfaces, Coatings and Films
Authors
, , , , , , , , , ,