Article ID Journal Published Year Pages File Type
1687218 Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 2017 4 Pages PDF
Abstract
The rearrangement allowed larger aperture diaphragms for higher beam currents at a moderate increase in beam spot sizes. Performance tests yielded proton microbeam currents and half-widths of 4.5 nA at 1.5 μm, 8.3 nA at 1.5 μm and 17.2 nA at 2 μm. For high resolution work the expected beam spots around 0.3 μm at 100 pA were not achieved. The reason is very likely interference on the beam scanner, correlated in x- and y-direction, which results from the insufficiently rectified power supply voltage of the transconductance amplifier.
Related Topics
Physical Sciences and Engineering Materials Science Surfaces, Coatings and Films