Abstract

A thorough knowledge of the primary spectrum is very important to perform a quality control (QC) of X-ray tubes. In previous works, a methodology to assess primary spectrum using a Compton spectrometer and applying the Monte Carlo (MC) method has been analyzed. By means of a Monte Carlo model, a response matrix can be built, relating the Pulse Height Distribution (PHD) recorded in the detector to the primary X-ray spectrum. Subsequently, an unfolding method based on the application of a Modified Truncated Singular Value Decomposition (MTSVD) is applied to the response matrix to assess the primary spectrum. Germanium (Ge), Silicon (Si) and Cadmium-Telluride (CdTe) detectors are considered in this work. The main goal of the paper is to determine whether the MTSVD unfolding method is adequate to provide an acceptable reproduction of spectra for these detectors.

► Ge, Si and CdTe detectors have been used with the MTSVD unfolding method. ► This methodology allows reconstructing primary spectra of X-ray. ► Spectra obtained with MTSVD agree with primary spectra provided by IPEM 78. ► MTSVD unfolding method could be used for quality control of the X-ray tubes in radiodiagnostic.