Article ID Journal Published Year Pages File Type
1759062 Ultrasonics 2009 7 Pages PDF
Abstract

This work characterizes the electrical and electromechanical aspects of an ultrasonic linear phased array inspection system, using a matrix of system functions that are obtained from the measured response of individual array elements in a simple reference experiment. It is shown that for the arrays tested all these system functions are essentially identical, allowing one to use a single system function to characterize the entire array, as done for an ordinary single element transducer. The variation of this single system function with the number of elements firing in the array or with changes of the delay law used is examined. It is also demonstrated that once such a single system function is obtained for an array, it can be used in a complete ultrasonic measurement model to accurately predict the array response measured from a reference reflector in an immersion setup.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Acoustics and Ultrasonics
Authors
, ,