Article ID Journal Published Year Pages File Type
1760215 Ultrasound in Medicine & Biology 2016 15 Pages PDF
Abstract
Model-based intravascular ultrasound elastography visualizes the stress distribution within vascular tissue-information that clinicians could use to predict the propensity of atherosclerotic plaque rupture. However, there are concerns that clusters of microvessels may reduce the accuracy of the estimated stress distribution. Consequently, we have developed a contrast-enhanced intravascular ultrasound system to investigate how plaque microvasculature affects the performance of model-based elastography. In simulations, diameters of 200, 400 and 800 μm were used, where the latter diameter represented a cluster of microvessels. In phantoms, we used a microvessel with a diameter of 750 μm. Peak stress errors of 3% and 38% were incurred in the fibrous cap when stress recovery was performed with and without a priori information about microvessel geometry. The results indicate that incorporating geometric information about plaque microvasculature obtained with contrast-enhanced ultrasound imaging improves the accuracy of estimates of the stress distribution within the fibrous cap precisely.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Acoustics and Ultrasonics