Article ID Journal Published Year Pages File Type
1760676 Ultrasound in Medicine & Biology 2014 10 Pages PDF
Abstract
This study assessed inter- and intra-observer reproducibility of shear wave elastography (SWE) measurements in vessel phantoms simulating soft and hard carotid plaque under steady and pulsatile flow conditions. Supersonic SWE was used to acquire cine-loop data and quantify Young's modulus in cryogel vessel phantoms. Data were acquired by two observers, each performing three repeat measurements. Mean Young's modulus was quantified within 2-mm regions of interest averaged across five frames and, depending on vessel model and observer, ranged from 28 to 240 kPa. The mean inter-frame coefficient of variation (CV) was 0.13 (range: 0.07-0.18) for observer 1 and 0.14 (range: 0.12-0.16) for observer 2, with mean intra-class correlation coefficients (ICCs) of 0.84 and 0.83, respectively. The mean inter-operator CV was 0.13 (range: 0.08-0.20), with a mean ICC of 0.76 (range: 0.69-0.82). Our findings indicate that SWE can quantify Young's modulus of carotid plaque phantoms with good reproducibility, even in the presence of pulsatile flow.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Acoustics and Ultrasonics
Authors
, , , , , ,