Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1761920 | Ultrasound in Medicine & Biology | 2011 | 15 Pages |
Abstract
We present a fast, automatic method for mitral annulus excursion measurement using pocket-sized ultrasound (PSU). The motivation is to provide PSU users with a rapid measurement of cardiac systolic function. The algorithm combines low-frame-rate tolerance, computational efficiency and automation in a novel way. The method uses a speckle-tracking scheme, initialized and constrained by a deformable model. A feasibility study using 30 apical four-chamber PSU recordings from an unselected patient population revealed an error of (mean ± SD) -1.80 ± 1.96 mm, p < 0.001, when compared with manual anatomic m-mode analysis using laptop scanner data. When only septal side excursion was measured, the mean error was -0.27 ± 1.89 mm, p < 0.001. The accuracy is comparable with previously reported results using semiautomatic methods and full-size scanners. The computation time of 3.7 ms/frame on a laptop computer suggests that a real-time implementation on a PSU device is feasible. (E-mail: sten.r.snare@ntnu.no)
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Acoustics and Ultrasonics
Authors
Sten Roar Snare, Ole Christian Mjølstad, Fredrik Orderud, Bjørn Olav Haugen, Hans Torp,