Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1784883 | Infrared Physics & Technology | 2010 | 8 Pages |
Abstract
A procedure is described where the response function of each pixel of an InSb detector is determined by radiometric self-calibration. With the present approach no knowledge of the spectral characteristics of the IR system is required to recover a quantity which is linear with the incident irradiance of the object. The inherent detector non-uniformity is corrected on the basis of self-calibrated scaled irradiance. Compared to the standard two-point non-uniformity correction procedure - performed with the detector signal - only two NUC-tables are required for arbitrary integration times. Images obtained at various exposures are fused to a single high dynamic range image. The procedure is validated with synthetic data and its performance is demonstrated by measurements performed with a high resolution InSb FPA.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Atomic and Molecular Physics, and Optics
Authors
M. Ochs, A. Schulz, H.-J. Bauer,