Article ID Journal Published Year Pages File Type
1803605 Journal of Magnetism and Magnetic Materials 2009 6 Pages PDF
Abstract

In this work, a simple technique to obtain the hysteresis loops of magnetic beads (Spherotech Inc.) in liquid suspension is presented. The magnetic measurements were taken in a DC Magnetic Property Measurement System (MPMS-SQUID sensor). Samples were based on ferromagnetic beads (surface-functionalized NH2, mean diameter 4.32 μm) prepared in three conditions: dry, suspended in sucrose solution and in suspension after functionalization with fluorophore. Special small containers (1.3 cm long) made of non magnetic plastic were designed to hold the beads in liquid. The results indicate that the bead's remnant magnetization is half of the value at maximum applied field in all cases. However, due to the additional degrees of rotational freedom, beads suspended in a liquid do not present coercivity. The use of ferromagnetic beads and magnetic elements of different architectures for applications in bioassays is also discussed.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Condensed Matter Physics
Authors
, , , , , , , , , , ,