Article ID Journal Published Year Pages File Type
1804218 Journal of Magnetism and Magnetic Materials 2007 5 Pages PDF
Abstract

We have studied the hybridization of target DNA in solution with probe DNA on magnetic beads immobilized on the channel sidewalls in a magnetic bead separator. The hybridization is carried out under a liquid flow and is diffusion limited. Two systems are compared: one with a straight microfluidic channel and one with an integrated staggered herringbone mixer. Fluorescence microscopy studies show that the hybridization is much more efficient in the system with the integrated mixer. The results, which are discussed in terms of a simple model, are relevant for any diffusion-limited reaction taking place on the surface in a microfluidic system.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Condensed Matter Physics