Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1822109 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2016 | 6 Pages |
Abstract
A fiducialization method based on vibrating wire is described dedicated to the problem of locating the magnetic center relative to external fiducials for the small-aperture quadrupoles. The advantage of this method is that the measurement of the wire position, which may be the main error source, is no longer needed. The position of the magnetic center can be directly obtained by measuring the position shift of the magnet fiducials. This method has been validated on small Permanent Magnet Quadrupoles (PMQs). Experiments have confirmed its feasibility of measuring PMQs with good repeatability of about 10 μm, and shown its high sensitivity as well as convenience.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Baichuan Wang, Shuxin Zheng, Lin Wu, Changtong Du, Qingzi Xing, Zhongming Wang, Mengtong Qiu, Xuewu Wang,