Article ID Journal Published Year Pages File Type
1822122 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2016 6 Pages PDF
Abstract

The characteristics and performance of a Dual Position Sensitive Multi-Wire Proportional Counter (DPS-MWPC) used to measure the scattering angle, the interaction position on the target and the velocity of reaction products detected in the VAMOS++ magnetic spectrometer, are reported. The detector consists of a pair of position sensitive low pressure MWPCs and provides both fast timing signals, along with the two-dimensional position coordinates required to define the trajectory of the reaction products. A time-of-flight resolution of 305(11)ps (FWHM) was measured. The measured resolutions (FWHM) were 2.5(3) mrad and 560(70)μm for the scattering angle and the interaction point at the target respectively. The subsequent improvement of the Doppler correction of the energy of the γ-rays, detected in the γ-ray tracking array AGATA in coincidence with isotopically identified ions in VAMOS++, is also discussed.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , , ,