Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1822338 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2015 | 7 Pages |
Abstract
In this paper we present thermal cycling experiments of GridPix radiation imaging detectors, in view of a potential application in a cryogenic experiment. The robustness of the GridPix detector is studied for various grid designs, as well as various mechanical and thermal surroundings. The grid design variations had insignificant effect on the grid strength. A low cool-down rate as well as good thermal contact are crucial for the durability of the grid. Further, additional strengthening at the grid edges proved necessary to maintain the integrity of the structure during thermal cycling, which was done using globtop adhesive. The combination of these measures led to 100% survival rate after thermal cycling down to −130 °C.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
R. Schön, J. Schmitz, S. Smits, Y. Bilevych, N. van Bakel,