Abstract

Recently, H3D has designed and introduced a gamma-ray imaging spectrometer system named Polaris-H. Polaris-H was designed to perform gamma spectroscopy and imaging throughout nuclear power plants. It integrates a 3D-position-sensitive pixelated CZT detector (20 mm×20 mm×15 mm), associated readout electronics, an embedded computer, a 5-h battery, and an optical camera in a portable water-proof enclosure. The total mass is about 4 kg, and the system startup time is 2 min. Additionally, it has a connection for a tablet, which displays a gamma-ray spectrum and isotope-specific images of the gamma-ray distribution in all directions in real time. List-mode data is saved to an external USB memory stick. Based on pixelated depth-sensing technology, spectroscopy is routinely better than 1.1% FWHM at 662 keV, and imaging efficiency at 662 keV varies less than a factor of two for all directions, except through the battery. Measurements have been performed in contaminated environments, in high radiation fields, and in cramped quarters.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , ,