Article ID Journal Published Year Pages File Type
1822524 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2015 8 Pages PDF
Abstract

This paper describes the design of the active muon veto subsystem for the MiniCLEAN dark matter direct detection experiment at SNOLAB in Sudbury, Ontario, Canada. The water-filled veto is instrumented with 48 PMTs which are read out by front-end electronics to time multiplex 48 photomultiplier channels into 6 digitizer channels and provide an instantaneous hit sum across the subsystem (N-Hit) for the veto trigger. We describe the primary system components: the PMTs, the support structure, the front-end electronics, and the data acquisition system.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , ,