Abstract
The time resolution of Multigap Resistive Plate Chambers (MRPCs) to 511 keV gamma rays has been investigated using a 22Na source and four detectors. The MRPCs time resolution has been derived from the Time-of-Flight information, measured from pairs of space correlated triggered events. A GEANT4 simulation has been performed to analyze possible setup contributions and to support experimental results. A time resolution (FWHM) of 312 ps and 376 ps has been measured for a single MRPC with four 250 μm gas gaps by considering respectively one and two independent pairs of detectors.These values, endorsed by the GEANT4 simulation, represent a good result compared to those reported in the literature.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
G. Belli, M. Gabusi, G. Musitelli, R. Nardò, S.P. Ratti, A. Tamborini, P. Vitulo,