Abstract

We present a straightforward method for particle identification and background rejection in 3He proportional counters for use in neutron detection. By measuring the risetime and pulse height of the preamplifier signals, one may define a region in the risetime versus pulse height space where the events are predominately from neutron interactions. For six proportional counters surveyed in a low-background environment, we demonstrate the ability to reject alpha-particle events with an efficiency of 99%. By applying the same method, we also show an effective rejection of microdischarge noise events that, when passed through a shaping amplifier, are indistinguishable from physical events in the counters. The primary application of this method is in measurements where the signal-to-background for counting neutrons is very low, such as in underground laboratories.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , ,