Abstract
► We developed a system for online dosimetry with RADFET sensors under zero bias. ► The system is calibrated for doses up to 10 kGy with REM RFT-300 sensors. ► We collected data on fade for over 300 days from irradiation. ► We present undulator dose measurements for 20 months of operation of FERMI@Elettra. ► Dose measurements are in good agreement with chemical film dosimeters.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation