Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1823390 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2013 | 13 Pages |
Abstract
We present some aspects of photon counting to study scintillators at low temperatures. A time-to-digital converter (TDC) had been configured to acquire several-minute-long streams of data, simplifying the multiple photon counting coincidence technique. Results in terms of light yield and time structure of a ZnWO4 scintillator are comparable to those obtained with a fast digitizer. Streaming data also provides flexibility in analyzing the data, in terms of coincidence window between the channels, and acquisition window of individual channels. We discuss the effect of changing these parameters, and use them to confirm low-energy features in the spectra of the number of detected photons, such as the 60Â keV line from 241Am in the ZnWO4 sample. We lastly use the TDC to study the transmission of the optical cryostat employed in these studies at various temperatures.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
P.C.F. Di Stefano, P. Nadeau, C.J.G. Onderwater, C. Trudeau, M.-A. Verdier,