Article ID Journal Published Year Pages File Type
1823482 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2012 9 Pages PDF
Abstract

We have developed and constructed a C-band (4.760 GHz) rf cavity beam position monitor (RF-BPM) system for the XFEL facility at SPring-8, SACLA. The demanded position resolution of the RF-BPM is less than 1μm, because an electron beam and x-rays must be overlapped within 4μm precision in the undulator section for sufficient FEL interaction between the electrons and x-rays. In total, 57 RF-BPMs, including IQ demodulators and high-speed waveform digitizers for signal processing, were produced and installed into SACLA. We evaluated the position resolutions of 20 RF-BPMs in the undulator section by using a 7 GeV electron beam having a 0.1 nC bunch charge. The position resolution was measured to be less than 0.6μm, which was sufficient for the XFEL lasing in the wavelength region of 0.1 nm, or shorter.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation