Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1823521 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2012 | 7 Pages |
Abstract
Over the past year, new 1 m×1 m neutron detectors have been installed at both the General Purpose SANS (GP-SANS) and the Bio-SANS instruments at HFIR, each intended as an upgrade to provide improved high rate capability. This paper presents the results of characterization studies performed in the detector test laboratory, including position resolution, linearity and background, as well as a preliminary look at high count rate performance.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Kevin D. Berry, Katherine M. Bailey, Justin Beal, Yacouba Diawara, Loren Funk, J. Steve Hicks, A.B. Jones, Kenneth C. Littrell, S.V. Pingali, P.R. Summers, Volker S. Urban, David H. Vandergriff, Nathan H. Johnson, Brandon J. Bradley,