A multi-reflection time-of-flight (MR-ToF) mass analyzer has been integrated into ISOLTRAP, the precision mass spectrometer for on-line mass determinations of short-lived nuclides at ISOLDE/CERN. The new instrument improves ISOLTRAP by providing a fast separation of isobaric contaminant species as well as subsequent ion selection using the fast Bradbury–Nielsen gate. Suppression ratios of up to 104 and mass-resolving powers of over 105 have been reached in off-line experiments. Preliminary data from on-line applications illustrate the benefit and performance of the device and its potential in the context of the ISOLTRAP setup.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
R.N. Wolf, D. Beck, K. Blaum, Ch. Böhm, Ch. Borgmann, M. Breitenfeldt, F. Herfurth, A. Herlert, M. Kowalska, S. Kreim, D. Lunney, S. Naimi, D. Neidherr, M. Rosenbusch, L. Schweikhard, J. Stanja, F. Wienholtz, K. Zuber,