Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1823818 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2012 | 6 Pages |
Abstract
A PET module that comprises a pair of compact, high-resolution and large-area detectors based on silicon photomultipliers (SiPM) was developed and evaluated. The detector's effective area was 24 mm×24 mm. LYSO crystals measuring 1.9 mm×1.9 mm×10 mm and charge division circuit were employed to obtain high spatial resolution with relatively concise Front-end electronics (FEE). Initial results showed the detector to be compatible with 1.5 T magnetic fields. The system's intrinsic FWHM spatial resolution is 1.50 mm. The average energy resolution is 18.5% and the coincidence timing resolution is 2.6 ns.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Yingjie Wang, Zhiming Zhang, Daowu Li, Baoyi Wang, Lei Shuai, Baotong Feng, Pei Chai, Shuangquan Liu, Haohui Tang, Ting Li, Yanfei Liao, Xianchao Huang, Yan Chen, Yantao Liu, Yiwen Zhang, Long Wei,