Article ID Journal Published Year Pages File Type
1823997 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2012 4 Pages PDF
Abstract

Characterization studies of the metal micro-detector TimePix measuring dose distribution at the Minibeam Radiation Therapy setup (Bio-Medical Beamline ID17, ESRF) were performed. The results obtained for high intensity synchrotron radiation minibeams illustrate an excellent performance of the TimePix providing in real time 2D image of the dose distribution over many beams in a 14×14 mm2 area. Peak-to-Valley–Dose–Ratios measured by TimePix and gafhromic films agree well.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , , ,