Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1824067 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2012 | 4 Pages |
Abstract
RPC timing properties are mainly determined by the maximum ionization rate that can be sustained while keeping the streamer fraction at modest values. The last characteristic, resilience to streamers, cannot be fully calculated, calling for an experimental approach. In this work we systematically study the influence of a range of gas mixtures on the performance (background counting rate, efficiency, time resolution and charge spectrum) of timing RPCs, using cosmic rays as ionizing particles.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
L. Lopes, P. Fonte, A. Mangiarotti,