A YAG(Ce) crystal has been characterized with a proton beam up to 100 MeV. Tests were performed to investigate the possibility of using this detector as a proton calorimeter. A crystal size has been chosen that is able to stop up to 200 MeV. Energy resolution and light response have been measured at Laboratori Nazionali del Sud with a proton beam up to 60 MeV and a spatial homogeneity study of the crystal has been performed at Loma Linda University Medical Center with a 100 MeV proton beam. The YAG(Ce) crystal showed a good energy resolution equal to 3.7% at 60 MeV and measurements, performed in the 30–60 MeV proton energy range, were fitted by Birks' equation. Using a silicon tracker to determine the particle entry point in the crystal, a spatial homogeneity value of 1.7% in the light response has been measured.