Article ID Journal Published Year Pages File Type
1824377 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2011 4 Pages PDF
Abstract

Qualitative measurements using digital neutron imaging has been the more explored aspect than accurate quantitative measurements. The reason for this bias is that quantitative measurements require correction for background and material scatter, and neutron spectral effects. Quantitative Neutron Imaging (QNI) software package has resulted from efforts at the Paul Scherrer Institute, Helmholtz Zentrum Berlin (HZB) and Necsa to correct for these effects, while the sample-detector distance (SDD) principle has previously been demonstrated as a measure to eliminate material scatter effect. This work evaluates the capabilities of the QNI software package to produce accurate quantitative results on specific characteristics of porous media, and its role to nondestructive quantification of material with and without calibration. The work further complements QNI abilities by the use of different SDDs. Studies of effective %porosity of mortar and attenuation coefficient of water using QNI and SDD principle are reported.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , ,