Abstract

Calibration factors w, for determination of fission rate in metallic foils of natU, 235U, 232Th, natPb and 197Au were determined for foils in contact with synthetic mica track detectors. Proton-induced fission at proton energies of 0.7 GeV and 1.5 GeV were used. Using our experimental results as well as those of the other authors, w for different foil–mica systems were determined. Two methods were used to calculate w, relative to the calibration factor for uranium–mica system, which has been obtained in a standard neutron field of energy 14.7 MeV. One of these methods requires the knowledge of the mean range of the fission fragments in the foils of interest and other method needs information on the values of the fission cross-sections at the required energies as well as the density of the tracks recorded in the track detectors in contact with the foil surfaces. The obtained w-values were compared with Monte Carlo calculations and good agreements were found. It is shown that a calibration factor obtained at low energy neutron induced fissions in uranium isotopes deviates only by less than 10% from those obtained at relativistic proton induced fissions.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , ,