Abstract
A full dosimetric characterization of the detector was performed with photon and proton beams and with gammas from a 60Co unit. Results have been compared with those obtained with a Farmer and a CC13 Scanditronix/Welhoffer ion chamber as well as with the silicon matrix MapCHECKâ¢. The first application of the MAESTRO prototype in the dosimetric verification of a clinical IMRT field is also reported. Results show that our modular detector represents a valuable tool for quality assurance in IMRT dose delivery and for high precision radiotherapy techniques.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
C. Talamonti, M. Bruzzi, L. Marrazzo, D. Menichelli, M. Scaringella, M. Bucciolini,