Article ID Journal Published Year Pages File Type
1825127 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2011 8 Pages PDF
Abstract

We have investigated by simulations and experimentally the parameters that affect image quality (contrast and spatial-resolution) of the fast neutron TRION detector. A scintillating fiber screen with 0.5×0.5 mm2 square fibers, few centimeters thick, provides superior spatial-resolution to that of a slab scintillator of the same thickness. A detailed calculation of the neutron interaction processes that influence the point-spread function (PSF) in the scintillating screen has been performed using the GEANT 3.21 code. The calculations showed that neutron scattering within the screen accounts for a significant loss of image contrast. The factors that limit the spatial-resolution of the image are the cross-sectional scintillating-fiber dimensions within the screen and the spatial response of the image-intensifier. A deconvolution method has been applied for restoring the contrast and the spatial-resolution of the fast neutron image.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , , ,